Membrane Residual Chlorine Meter (online transmitter)
SKU: WQ-00358
Applications
It is widely used for continuous monitoring and control of residual chlorine levels in water solutions in various applications, such as drinking water treatment plants, water distribution networks, swimming pools, cooling water systems, and water quality treatment projects. 96x96 intelligent online residual chlorine transmitter. Multi-parameter display on the same screen, residual chlorine, hypochlorous acid, temperature, pH value, output current. Restore factory setting function, slope correction function. RS485 MODBUS standard communication protocol. One way isolated 4~20mA output, programmable corresponding to any residual chlorine value. CL-1001.
- Measuring range
- Residual chlorine: 0~20mg/L
- HOCL
- 0~10.00mg/L
- PH
- 0~14.00PH
- Temp
- 0~60℃
- Resolution
- Residual chlorine: 0.01mg/L
- HOCL
- 0.01mg/L
- PH
- 0.01PH
- Temp
- 0.1℃
- Accuracy
- Residual chlorine: ±1% or ±0.01mg/L
- HOCL
- ±1% or ±0.01mg/L
- pH
- ±0.02PH
- Temp
- ±0.5℃
- Compensation
- Manual/automatic pH compensation function(0-14) and temperature compensation(0~60℃)
- Operating temperature
- 0℃~60℃, RH≤90%
- Storage temperature
- -20℃~+70℃
- Input impedance
- >1012Ω
- Screen display
- Back light type large LCM display
- Analog output
- One way isolated 4~20mA output
- Digital output
- RS485 Modbus standard communication protocol.
- Relay output
- High/low points and delay are programmable
- Power supply
- 90~240V AC, 50/60Hz
- Protection level
- IP65
- Installation
- Wall, panel, pipe
- Overall size(HxWxD)
- 96mm x 96mm x 130mm
- Hole size(H x W)
- 93mm x 93mm
- Weight
- 0.5KG